Product list

Contact us

Contact:Manager Liang/Manager Qin
Postcode:528305
Add:Ronggui shunde district, Foshan city, Guangdong province, streets GuiFu road no. 36 ,3 street
Tel:0757-28873576 / 0757-28991293 / 0757-28991296
Fax:0757-28872395 或(010)950507转727958
E_mail:sdkyd@263.net

Dialyzer Auto-reprocessing machine/system

position : Home - Dialyzer Auto-reprocessing machine/system

KYD-688A Dialyzer Auto-reprocessing system

 

 

  

 

 

     Automatic processor/system maintenance (multiplexing machine/system) is a kind of 6845 extracorporeal circulation and blood treatment for chronic hemodialysis (Hemodialysis auxiliary equipment) purpose: auxiliary equipment of hollow fiber maintenance blood automatically rinse,rinse membrane complete detection,capacity test,inject the disinfectant,print label ,to maintenance of reusable requirements.

 

Function Configuration

 Function Configuration

 688 Dialyzer Auto-reprocessing System( A )

 1

Dispose dialyzer quantity once

 2

structure 

host machine(1)、disposer (1)

 3

Color LCD,touch screen display

 √

 4

Dispose once time   

(10~13)min/1 only

 5

Blood chamber (film) is in flushing

 √

 6

Blood chamber (film) in reverse wash

 √

 7

Dialysis fluid chamber (film) is the flushing  

 √

 8

Dialysis fluid chamber (film) in reverse flushing  

 √

 9

Pulse overhaul filter flush

 √

 10

Flush flow

average>4L/mim(peak 6L/min)

 11

Innovative constant pressure flush device

 √

 12

Adjustable reverse-ultrafiltration pressure to adapt to various transmembrane pressure

 (0.5~2)Kg/cm2

 13

Disinfection liquid injection pattern

 Compatible with various chemical to disinfect dialyzer innovative metering pump withmagnetci drive accur-tely fill chemical

 14

Automatic dilution any brand original disinfection  

Optional external automatic dilution system

 15

Machine wash

 √

 16

Machine disinfection 

 √

 17

Auto detect membrane integrity  

The blood compartment is tested at a positive 250mmHg.A decay in pressure at a rate equal to or less than 30mmhg per 30 seconds

 18

Capacity test 

range:0~170ML,Measured value error±3%

 19

Alarm function

Water shortages、Lack of disinfection、Pressure test failure

 20

Prompt functions

Rinse finish、Sanitize Cycle complete、ressure test pass

 21

dialyzer database

Can reprocess all kinds hollow fiber dialyzers,system can detect and match the dialyzer,also manual input and match the dialyzer various

 22

Material input methods 

MMI

  √

code

   -

write

  √

scanistor

 match

 23

patient Database

 $forty thousand processing information storage, search, print

 24

Built-in printer

Name tag, testing label printing

 25

Professional maintenance interface 

  √

 26

Power On Self Test 

  √

 27

Automatic processing

 Automatic or manual process;Manual Process consist of four individual process,each process can be done step by step

 28

To water pan

 match

 29

Case size

 high49cm、wide36cm、long45.5cm

 

 Use conditions

Power supply

 220V.AC/50Hz

Power

 <800VA

Work environment

 temperature:5℃~40℃;

Running time

 (10~12)min/1 only

Water quality requirements

 The system comply to the Dialyzer Reprocessing Guideline of chinese MOH

Supply pressure

200~250kPa 

Water flow

 (3-6)L/min

Water temperature

 10℃~38℃

Drainage requirements

 Drainage function>6L/min